C. diff-CUBE, a dot immunobinding assay (DIA) (Difco Laboratories, Ann Arbor, Mich.) for detection of Clostridium difficile toxin A in stool specimens, was compared with latex agglutination (LA) (Marion Laboratories, Kansas City, Mo.) and cytotoxin assay (CTA) for the laboratory diagnosis of C. difficile-associated diarrhea. A total of 200 stool specimens collected from 169 patients with suspected C. difficile diarrhea were tested. Of the 198 specimens evaluated by all three methods, 36 (18%) from 36 patients were positive by one or more of the tests. Twenty-five, 26, and 23 specimens were positive by CTA, DIA, and LA, respectively; 14 were positive by all three methods. Eight specimens yielded nonspecific LA test results; all eight were negative by CTA, and one was positive by DIA. DIA results agreed with CTA results in 183 (92%) cases and with LA results in 175 (88%) cases. CTA and LA results agreed in 179 (90%) cases. Freezing of the specimen did not appear to adversely affect either the DIA or LA test. These preliminary results suggest that C. diff-CUBE may be useful as a rapid screen for the diagnosis of C. difficile-associated diarrhea. However, for optimum laboratory diagnosis, further testing of all stools that are negative by DIA is warranted.
No takes yet. Share an insight, caveat, or question.
Woods et al. (1990) studied this question.