Why the study?
T and small protrusion (TAP) stenting is a suboptimal technique for bailout side branch stenting in coronary bifurcation lesions.
Does modified TAP (mTAP) stenting improve stent area and reduce neocarina length compared to standard TAP stenting in coronary bifurcation lesions?
Does modified TAP (mTAP) stenting improve stent area and reduce neocarina length compared to standard TAP stenting in coronary bifurcation lesions?
A novel modified TAP stenting technique for coronary bifurcation lesions demonstrated improved stent area and shorter neocarina length in bench testing, with favorable initial imaging and clinical outcomes in a small patient cohort.
Answers are grounded in this paper and cite their sources.
mTAP may optimize bailout SB stenting in CBL; hypothesis-generating and requires prospective randomized validation before practice change.
Objectives We investigated the results of T and small protrusion (TAP) versus a novel modification of TAP (mTAP) stenting by multimodality imaging in bench testing and in patients with coronary bifurcation lesions (CBL). Background TAP stenting is a suboptimal technique for bailout side branch (SB) stenting. Methods In a bench model, optical coherence tomography (OCT), microscopic examinations (ME), and computational fluid dynamics (CFD) were performed after TAP and mTAP stenting. In 20 patients with CBL, 80 intravascular ultrasound (IVUS) examinations were performed during mTAP stenting in which the SB stent was pulled‐back to indent the inflated main vessel (MV) balloon and deployed while deflating it. For TAP stenting, the tip of the SB stent was positioned in the MV and deployed. Results In bench testing, OCT showed neocarina length (NL) was shorter and minimum stent area (MSA) was larger after mTAP versus TAP stenting (2.84 ± 0.70 vs. 4.80 ± 020 mm; 6.75 ± 1.50 vs. 4.5 ± 2.2 mm 2 ; respectively; p < 0.05). By ME, NL was shorter and shear rate trended lower after mTAP versus TAP stenting. In patients, IVUS showed MSA was larger after versus before mTAP stenting (6.32 ± 0.58 vs. 5.21 ± 0.56 mm 2 ; p < 0.01); NL was 1.43 ± 0.22 mm with SB ostium coverage. The Seattle Angina questionnaire (SAQ) score was higher at 6 months versus baseline (85 ± 4.0 vs. 48 ± 6.0, respectively; p < 0.001). Conclusions This multimodality imaging study showed, for the first time, mTAP stenting resulted in larger stent area and shorter neocarina than TAP stenting in bench testing. In patients with CBL, mTAP stenting led to larger stent area, short neocarina with complete SB ostium coverage, and improved the SAQ score at follow‐up.
No takes yet. Share an insight, caveat, or question.
Ng et al. (2022) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: