A glucose electrode was constructed by adsorbing glucose oxidase (GOD) on a modified electrode for H2 O 2 oxidation, consisting of Pd/Au sputtered on graphite. Maximally, 0.8 U cm−2 of GOD could be adsorbed. The electrode was used in a f.i.a. manifold for determination of glucose. Linear calibration curves were obtained in the concentration range 3. 10−6 4. 10−3 mol L−1 glucose. The applied potentials for glucose determination were + 300 mV vs. Ag/AgCl at pH 8.0, + 350 mV at pH 7.0, + 400 mV at pH 6.0 and + 500 mV at pH 5.0. The activity vs. pH profile of adsorbed GOD was broad having an optimum between pH 5 and 6. The apparent kinetic parameters for adsorbed GOD, KM app and imax, were found to be 50 mM and 160 uA at optimal pH.
No takes yet. Share an insight, caveat, or question.
Jönsson et al. (1987) studied this question.
Synapse has enriched 5 related papers. Consider them for comparative context: