In this paper, a time-domain system dedicated to medical diagnostics has been designed, a prototype has been built and its performance has been evaluated. Measurements show that the system has a 3-dB bandwith of about 3.5 GHz and a signal to noise ratio over 40 dB in the frequency range about 800 MHz to 3.8 GHz. The system has been used to perform a microwave tomographic image reconstruction test. The same target was reconstructed based on data measured with a network analyzer. A comparison between the images shows very small differences, and proves the functionality of the time domain system.
No takes yet. Share an insight, caveat, or question.
Zeng et al. (2014) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: