Why the study?
Telehealth ECGs are noisier than traditional clinical ECGs, presenting challenges for automated QRS detection, and the best-performing open-source detector for telehealth ECGs remained unidentified.
Do open-source QRS detection algorithms accurately detect QRS complexes in single-lead telehealth ECG signals compared to manual annotations?
Do open-source QRS detection algorithms accurately detect QRS complexes in single-lead telehealth ECG signals compared to manual annotations?
Top-performing open-source QRS detectors (Neurokit, two-avg, UNSW) are highly accurate on high-quality telehealth ECGs but struggle with low-quality signals, emphasizing the need for automated signal quality assessment in remote monitoring.
Caution advised applying standard QRS detectors to noisy telehealth ECGs; leaves open optimal open-source choice pending validation.
Background and Objectives A key step in electrocardiogram (ECG) analysis is the detection of QRS complexes, particularly for arrhythmia detection. Telehealth ECGs present a new challenge for automated analysis as they are noisier than traditional clinical ECGs. The aim of this study was to identify the best-performing open-source QRS detector for use with telehealth ECGs. Methods The performance of 18 open-source QRS detectors was assessed on six datasets. These included four datasets of ECGs collected under supervision, and two datasets of telehealth ECGs collected without clinical supervision. The telehealth ECGs, consisting of single-lead ECGs recorded between the hands, included a novel dataset of 479 ECGs collected in the SAFER study of screening for atrial fibrillation (AF). Performance was assessed against manual annotations. Results A total of 12 QRS detectors performed well on ECGs collected under clinical supervision ( F 1 score ≥ 0.96). However, fewer performed well on telehealth ECGs: five performed well on the TELE ECG Database; six performed well on high-quality SAFER data; and performance was poorer on low-quality SAFER data (three QRS detectors achieved F 1 of 0.78-0.84). The presence of AF had little impact on performance. Conclusions The Neurokit and University of New South Wales QRS detectors performed best in this study. These performed sufficiently well on high-quality telehealth ECGs, but not on low-quality ECGs. This demonstrates the need to handle low-quality ECGs appropriately to ensure only ECGs which can be accurately analysed are used for clinical decision making.
No takes yet. Share an insight, caveat, or question.