Answers are grounded in this paper and cite their sources.
Objective To verify the feasibility of geostationary orbit satellite communication in ultra-remote robotic surgery and to construct a remote surgical operation control technology system under high-latency communication conditions. Methods A large time-delay control method was proposed, including the establishment of the Stability Operation Criteria for Large Time-Delay Remote Robotic Surgery and Generalized-Order Compliant Interaction Control Method. On December 26, 2024, a transregional communication link between Lhasa and Beijing was established via the Asia-Pacific 6D high-throughput satellite (orbital altitude of 36,000 km), and two hepatectomies for liver cancer were performed using the Tumai surgical robot system. Core parameters such as end-to-end latency and data packet loss rate were monitored. Results The Lhasa-Beijing link established through the Asia-Pacific 6D satellite achieved an average end-to-end latency of 632 ms (measured uplink 4.2 Mbps/downlink 3 Mbps), with intraoperative robotic arm pose tracking error <0.5 mm (n=2). Both hepatectomies for liver cancer were successfully completed, with surgery durations of 115-124 min, blood loss of 20 mL, and patients discharged within 24 h post-surgery without severe complications (Clavien-Dindo I level). Conclusion This study is the first to confirm the safety of geostationary orbit satellite support for human surgery and verifies the feasibility of surgical operation control technology under high-latency conditions.
No takes yet. Share an insight, caveat, or question.
Zhang et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: