Why the study?
Carillon device efficacy in reducing FMR might depend on the anatomical relation between the coronary sinus and the mitral annulus.
Does CT-angiography assessment of the distance and angle between the coronary sinus and mitral annulus predict reduction in functional mitral regurgitation after Carillon device implantation in patients with FMR?
Does CT-angiography assessment of the distance and angle between the coronary sinus and mitral annulus predict reduction in functional mitral regurgitation after Carillon device implantation in patients with FMR?
Pre-procedural CT-angiography assessment of the distance and angle between the coronary sinus and mitral annulus can help predict echocardiographic response to indirect mitral annuloplasty using the Carillon device.
Answers are grounded in this paper and cite their sources.
CT assessment of CS-mitral annulus relation may refine Carillon selection; leaves open prospective validation of anatomy-based response prediction.
OBJECTIVES: Coronary sinus (CS) based mitral annuloplasty using the Carillon device is effective in reducing functional mitral valve regurgitation (FMR). However, this positive effect might be dependent on the relation between CS and the mitral annulus. BACKGROUND: Computed tomography (CT) assessment prior to mitral valve interventions is an emerging technique to optimize patient selection. METHODS: In a retrospective analysis 30 patients underwent Carillon device implantation and received CT-angiography prior to CS based percutaneous mitral valve repair. Patients were assigned to responders or non-responders according to the 3-month transthoracic echocardiographic follow-up including quantitative mitral valve regurgitation assessment. A prototype software for CS reconstruction was used to assess distance and angle of both CS and mitral annulus planes. RESULTS: Comparison of the distance and angle of the CS plane and the mitral valve annulus plane showed a significant shorter distance and lower angle in the responder group implicating an impact on procedure success. Our results suggest a CS plane and MV annulus plane with a favorably distance of <7.8 mm and an optimal angle of <14.2° could be considered favorably for mitral annuloplasty using a Carillon device. CONCLUSIONS: Distance and angle of mitral annulus and CS planes determined by three-dimensional reconstructions of CT-angiography might predict a reduction in echocardiographic FMR using Carillon Mitral Contour System.
No takes yet. Share an insight, caveat, or question.
Rottländer et al. (2020) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: