PURPOSE: The purpose of this study was to assess the image quality of color K-edge imaging obtained with a spectral photon-counting CT (SPCCT) scanner using a spectral phantom with a mixture of iodine-based and gadolinium-based contrast agents. MATERIALS AND METHODS: A clinical SPCCT scanner prototype was used to scan a spectral phantom. Three dedicated cavities were filled with three contrast agents including iodine alone, gadolinium alone and a mixture of both. Two concentrations of 0.5 and 2 mg/mL were evaluated using nine helical PCCT scans at 120 kVp and 150 mAs. Conventional, color iodine and color K-edge gadolinium images were obtained through a material decomposition algorithm using three basis materials (water, iodine, gadolinium). Attenuation (in Hounsfield unit [HU]), iodine and gadolinium concentrations and task-based transfer function (TTF) were measured on each cavity and image. The noise power spectrum (NPS) was calculated on the phantom's background. The value of noise magnitude was 24.75 HU, 0.06 mg/mL and 0.03 mg/mL for conventional, color iodine and color K-edge gadolinium images, respectively. CONCLUSION: Color K-edge imaging helps distinguish between contrast agents while being associated with lownoise magnitude, high-frequency spatial noise and high spatial resolution.