Background The electrocardiogram (ECG) is important for gating purposes in cardiac magnetic resonance imaging (CMR). However, the magnetohydrodynamic (MHD) effect, which is caused by the flow of blood in the static magnetic field, makes it difficult to record clean ECG signals for gating. The vectorcardiogram (VCG), which can be derived from the ECG signal, is commonly used for gating purposes and has been well established for magnetic fields strength of up to 3T. However, for higher field strengths this method is prone to errors [1]. This work tries to explain the reasons for the VCG based methods not being suitable for cardiac gating in ultra high field MRI. Methods ECGs were recorded using a standard 12-lead Holter ECG (Getemed, Germany) inside a 7T MR scanner (Siemens Magnetom), a 3T MR scanner (Philips Achieva) and outside the MR scanner as references. Measurements were made on five healthy volunteers while MR imaging was switched off. The VCGs were derived from the 12-lead ECG using the inverse Dower matrix [2]. The VCG method [3] was implemented and applied to the acquired data sets. This method measures the distance in the VCG space between a static reference vector defined by the R-wave (R) peak of an ECG signal recorded outside the MR scanner (at 0T) and the VCG vector recorded inside the MR scanner. Those points where a minimal distance and angle between both vectors is reached are classified as R peaks.