A quartz crystal microbalance (QCM) spray-coated with a Ni(SCN)(2)(4-picoline)(4) film is a sensitive detector for small aromatic (benzene, toluene) and chlorinated (trichloroethylene, perchloroethylene) vapors with a planar molecular geometry. Frequency changes during transient exposures to these vapors are rapid and reversible. In contrast, frequency changes during transient exposures to carbon tetrachloride vapor exhibit a very slow rise and decay. Impedance studies demonstrate that the QCMs are responding only to mass changes in the film. Calibration curves exhibit both linear and near-saturation responses, depending on the vapor and vapor concentration. Partition coefficients obtained from the linear response regimes of the calibration curves are in the 10 000-100 000 range, more than an order of magnitude larger than the partition coefficients for a prototypical soft polymer, poly(isobutylene). Despite the absence of evidence for crystallinity by optical or X-ray diffraction methods, the spray-coated films appear to be forming clathrates with the organic vapors. The Ni(SCN)(2)(4-picoline)(4) film is promising for the development of very sensitive and partially selective piezoelectric sensors for nonpolar or weakly polar organic vapors in air.
No takes yet. Share an insight, caveat, or question.
Finklea et al. (1998) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: