BACKGROUND: We aimed to investigate the performance of the Space GlucoseControl system (SGC) (B. Braun, Melsungen, Germany) in medical critically ill patients. The SGC is a nurse-driven, computer-assisted device for glycemic control combining infusion pumps with the enhanced Model Predictive Control algorithm. SUBJECTS AND METHODS: The trial was designed as a single-center, open clinical investigation in a nine-bed medical intensive care unit in a tertiary center in Graz, Austria. Efficacy was assessed by percentage of time within the target range (4.4-8.3 mmol/L; primary end point), mean blood glucose, and sampling interval. Safety was assessed by the number of hypoglycemic episodes (≤2.2 mmol/L). RESULTS: Twenty mechanically ventilated patients (age, 63±16 years; body mass index, 31.0±10.7 kg/m(2); Acute Physiology and Chronic Health Evaluation II score, 25.4±6.3; 14 males; six with diabetes) were included for a period of 7.0±3.6 days. Time within target range was 83.4±8.9% (mean±SD), and mean arterial blood glucose was 6.8±0.4 mmol/L. No severe hypoglycemic episodes (<2.2 mmol/L) occurred, and the percentage of time within 2.2 and 3.3 mmol/L was low (0.03±0.07%). The sampling interval was 2.0±0.4 h. The mean insulin dose was 93.5±80.1 IU/day, and the adherence to the given insulin dose advice was high (98.3%). A total of 11 unintended therapy interruptions (0.08 events/treatment day) caused by software problems occurred in four patients. CONCLUSIONS: SGC is a safe and efficient method to control blood glucose in critically ill patients in the medical intensive care unit.
No takes yet. Share an insight, caveat, or question.
Amrein et al. (2012) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: