Key result
Transdermal optical imaging accurately measured heart rate and basal stress, demonstrating strong positive correlations (r=1.00 and r=0.89, respectively) with the standard BIOPAC ECG system.
Why the study?
Does transdermal optical imaging accurately measure heart rate and heart rate variability compared to standard electrocardiography in healthy adults?
Cross-Sectional (n=136)
Does transdermal optical imaging accurately measure heart rate and heart rate variability compared to standard electrocardiography in healthy adults?
Transdermal optical imaging is a valid, contactless, and remote methodology that can accurately measure heart rate and heart rate variability to assess basal stress, performing comparably to standard electrocardiography.
Supports contactless HR monitoring in research; leaves open clinical validation for HRV and stress assessment.
The present study examined the validity of a novel physiological measurement technology called transdermal optical imaging (TOI) technology at assessing basal stress. This technology conveniently, contactlessly, and remotely measures facial blood flow changes using a conventional digital video camera. We compared data from TOI against the pulse data collected from the FDA approved BIOPAC system. One hundred thirty-six healthy adults participated in the study. We found that TOI measurements of heart rate and heart rate variability (HRV), which reflects basal stress, corresponded strongly to those obtained from BIOPAC. These findings indicate that TOI technology is a viable method to monitor heart rate and HRV not only accurately but also conveniently, contactlessly, and remotely. Further, measures of HRV obtained via TOI serves as a valid index of basal stress. Potential applications of this technology in psychological research and other fields are discussed.
No takes yet. Share an insight, caveat, or question.
Hong et al. (2018) conducted a cross-sectional in Healthy adults (n=136). Transdermal optical imaging (TOI) vs. BIOPAC ECG system was evaluated on Correlation between TOI and BIOPAC measurements for heart rate and stress (SD1/SD2) (r=1.00 for heart rate, r=0.89 for stress). Transdermal optical imaging accurately measured heart rate and basal stress, demonstrating strong positive correlations (r=1.00 and r=0.89, respectively) with the standard BIOPAC ECG system.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: