This is the authors' abstract. We don't add key points for this paper.
The performance of a cryogenically cooled double-crystal silicon monochromator was studied under high-heat-load conditions with total absorbed powers and power densities ranging from 8 to 780 W and from 8 to 240 W mm(-2), respectively. When the temperature of the first crystal is maintained close to the temperature of zero thermal expansion of silicon, the monochromator shows nearly ideal performance with a thermal slope error of 0.6 µrad. By tuning the size of the first slit, the regime of the ideal performance can be maintained over a wide range of heat loads, i.e. from power densities of 110 W mm(-2) (at total absorbed power of 510 W) to 240 W mm(-2) (at total absorbed power of 240 W).
No takes yet. Share an insight, caveat, or question.
Chumakov et al. (2014) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: