Why the study?
Does short-time-window Patlak imaging using a population-based input function and BPL reconstruction provide comparable image quality and quantitative parameters to long-time-window imaging with individual patient-based input function in patients with cardiac sarcoidosis?
Does short-time-window Patlak imaging using a population-based input function and BPL reconstruction provide comparable image quality and quantitative parameters to long-time-window imaging with individual patient-based input function in patients with cardiac sarcoidosis?
Short-time-window Ki imaging using a population-based input function and BPL reconstruction with a high beta value is a feasible alternative to long-time-window imaging with individual patient-based input function for cardiac sarcoidosis.
May support shorter Ki imaging protocols in cardiac sarcoidosis; leaves open prospective validation before clinical adoption.
BACKGROUND: F-FDG-PET/CT scans of 30 patients with cardiac sarcoidosis. For each dynamic scan, the Ki images were obtained using the IF from each individual patient and a long time window (10-60 min). In addition, Ki images were obtained using the normalized averaged population-based IF and BPL algorithms with different beta values (350, 700, and 1000) with a short time window (40-60 min). The visual quality of each image was visually rated using a 4-point scale (0, not visible; 1, poor; 2, moderate; and 3, good), and the Ki parameters (Ki-max, Ki-mean, Ki-volume) of positive myocardial lesions were measured independently by two readers. Wilcoxon's rank sum test, McNemar's test, or linear regression analysis were performed to assess the differences or relationships between two quantitative variables. RESULTS: Both readers similarly rated 51 scans as positive (scores = 1-3) and 22 scans as negative (score = 0) for all four Ki images. Among the three types of population-based IF Ki images, the proportion of images with scores of 3 was highest with a beta of 1000 (78.4 and 72.5%, respectively) and lowest with a beta of 350 (33.3 and 23.5%) for both readers (all p < 0.001). The coefficients of determination between the Ki parameters obtained with the individual patient-based IF and those obtained with the population-based IF were highest with a beta of 1000 for both readers (Ki-max, 0.91 and 0.92, respectively; Ki-mean, 0.91 and 0.92, respectively; Ki-volume, 0.75 and 0.60, respectively; and all p < 0.001). CONCLUSIONS: Short-time-window Ki images with a population-based IF reconstructed using the BPL algorithm and a high beta value were closely correlated with long-time-window Ki images generated with an individual patient-based IF. Short-time-window Ki images using a population-based IF and BPL reconstruction might represent practical alternatives to long-time-window Ki images generated using an individual patient-based IF.
No takes yet. Share an insight, caveat, or question.
Tanaka et al. (2022) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: