A simple and high throughput method was developed and validated for simultaneous determination of valproic acid and its two toxicant ene-metabolites, 2-enevalproic acid and 4-enevalproic acid in epilepsy patient plasma using liquid chromatography–tandem mass spectrometry. Probenecid was used as internal standard and solid-phase extraction was selected for sample preparation. A chromatographic separation was performed on an Agilent Poroshell SB-C 18 column (50 mm×4.6 mm i.d., 2.7 μm) by an optimized gradient elution at a flow rate of 0.9 mL/min. The total run time was 7 min. Electrospray ionization was used in negative ion mode by multiple reaction monitoring of the precursor-to-product ion transitions at m / z 143.0→143.0 for valproic acid, m / z 140.9→140.9 for 2-enevalproic acid and 4-enevalproic acid for their poor fragments, and m / z 283.9→239.9 for probenecid. The results showed good linearity of valproic acid, 2-enevalproic acid and 4-enevalproic acid in their respective linear ranges. The correlation coefficients were more than 0.998. The intra- and inter-day precision of the assay was less than 11.0% and the accuracy ranged from 2% to 12%. This analytical method was successfully applied to assay plasma concentrations of valproic acid and its two ene-metabolites in epilepsy patient plasma and used for therapeutic drug monitoring.