Dither demodulation of fiber Bragg grating sensors illuminated with multimode light from laser diodes is theoretically and experimentally investigated. Quasi-static temperature and strain sensitivities of 0.09 degrees C/ radical Hz and 0.6 microepsilon/ radical Hz are obtained. We show that it is possible to measure small ac signals that lie outside the feedback loop bandwidth by using a synchronous detection referenced to twice the dither frequency. In this situation, dynamic strain sensitivity of 3.3 n(epsilon)/ radical Hz is achieved.