Key result
Addition of 20 µl PGE1 to the Multiplate ADP assay significantly reduced Arbitrary Aggregation Units but did not improve its moderate correlation with the VerifyNow assay.
Why the study?
Platelet function tests like Multiplate and VerifyNow show little concordance, which may be influenced by VerifyNow using prostaglandin E1 or differences in baseline platelet function.
Does the addition of prostaglandin E1 (PGE1) or the use of inhibitor percentage improve the correlation between Multiplate and VerifyNow platelet function tests in patients on dual antiplatelet therapy after PCI?
Cross-Sectional (n=51)
Does the addition of prostaglandin E1 (PGE1) or the use of inhibitor percentage improve the correlation between Multiplate and VerifyNow platelet function tests in patients on dual antiplatelet therapy after PCI?
Neither the addition of prostaglandin E1 nor the use of inhibitor percentage improves the moderate correlation between Multiplate and VerifyNow platelet function tests in patients on dual antiplatelet therapy.
Should not treat Multiplate and VerifyNow as equivalent; challenges PGE1 as the discordance driver and leaves open other technical factors.
Platelet function tests (PFT), such as the Multiple Electrode Analyzer (Multiplate) and VerifyNow, show little concordance in patients using antiplatelet drugs. A major difference between these tests is the use of prostaglandin E1 (PGE1) to inhibit P2Y1-platelet-receptor activation in VerifyNow and is proposed to be of influence in the discrepancy between these tests. We aimed to investigate whether the presence of PGE1 could provide an explanation for the moderate correlation and concordance between Multiplate and VerifyNow by adding PGE1 to the Multiplate ADP assay, also known as the ADP-high sensitivity (ADP-HS) assay. We also aimed to investigate whether the difference in baseline platelet function as measured by the VerifyNow and Multiplate could (partly) explain the moderate correlation between the tests, by plotting ADP assay results against baseline function as measured by the corresponding device, which is expressed as the 'inhibitor percentage.' Fifty-one patients who underwent percutaneous coronary intervention (PCI) received dual antiplatelet therapy and were considered to have a high risk of ischemic or bleeding complications were included. The addition of 20 µl PGE1 in the Multiplate resulted in a significant reduction in Arbitrary Aggregation Units, but did not improve correlation with the VerifyNow. The correlation between VerifyNow and Multiplate inhibitor percentage was moderate. Based on these results, we concluded that neither PGE1 nor the calculation of the inhibitor percentage greatly influenced the correlation between PFTs.
No takes yet. Share an insight, caveat, or question.
Hulshof et al. (2020) conducted a cross-sectional in Patients on dual antiplatelet therapy post-PCI at high risk of complications (n=51). Addition of prostaglandin E1 (PGE1) to Multiplate ADP assay vs. Standard Multiplate ADP assay and VerifyNow was evaluated on Correlation between Multiplate and VerifyNow assays. Addition of 20 µl PGE1 to the Multiplate ADP assay significantly reduced Arbitrary Aggregation Units but did not improve its moderate correlation with the VerifyNow assay.