Cardiac implantable electronic devices can cause artifacts in late gadolinium enhancement cardiac magnetic resonance imaging, and no consensus exists on optimal patient selection.
Does device type and generator-to-lead distance predict LGE-CMR image quality in patients with cardiac implantable electronic devices?
Does device type and generator-to-lead distance predict LGE-CMR image quality in patients with cardiac implantable electronic devices?
Device type (ICD vs PM) and generator-to-lead distance are significant predictors of LGE-CMR image quality in patients with CIEDs, providing practical thresholds to optimize pre-scan patient selection.
May guide CIED patient selection for LGE-CMR by device type and distance; leaves open prospective validation before practice change.
Background Cardiac implantable electronic devices (CIEDs) can cause artifacts in late gadolinium enhancement (LGE) cardiac magnetic resonance imaging (CMR), compromising diagnostic accuracy. No consensus exists on optimal CIED-patient selection for LGE-CMR. Objective This study aims to identify predictors of LGE image quality in CIED-patients to optimize pre-scan selection. Methods Patients with CIEDs who underwent conventional 2D-LGE imaging were retrospectively identified from the Amsterdam UMC CMR database. Baseline clinical and device characteristics were collected, and generator-to-lead distance was measured on post implantation chest X-rays. LGE-quality was categorized as fully diagnostic, acceptable, or non-diagnostic. Multivariable regression and ROC analysis determined independent predictors and exploratory generator-to-lead distance thresholds using a 90% sensitivity criterion. Results Eighty patients (71.3% male, mean age 64 years) were included: 41.3% ICDs, 23.8% PMs, 23.8% CRT-Ds, and 2.5% CRT-Ps. LGE image quality was fully diagnostic in 48.8%, acceptable in 27.5%, and non-diagnostic in 23.8%. PM/CRT-P patients had no non-diagnostic scans (92.9% fully diagnostic). Only 25.0% of ICD/CRT-D scans were fully diagnostic, while 36.5% were non-diagnostic (p < 0.001). Generator-to-lead distance was significantly associated with LGE-quality, with thresholds of 10cm in ICDs and 8cm in PMs for acceptable LGE-quality. Conclusion Device type and positioning significantly impact LGE-image quality. ICDs were associated with poorer image quality, while PMs consistently yielded diagnostic-quality images. Generator-to-lead distance emerged as a key predictor, providing a practical tool for optimizing LGE-CMR referrals. This study defines generator-to-lead distance thresholds by device type and proposes a structured pre-scan workflow to support LGE-CMR referral decisions in patients with CIEDs.
No takes yet. Share an insight, caveat, or question.
Hopman et al. (2025) studied this question.