At-wavelength interferometric characterization of a new 4×-reduction lithographic-quality extreme ultraviolet (EUV) optical system is described. This state-of-the-art projection optic was fabricated for installation in the EUV lithography Engineering Test Stand (ETS) and is referred to as the ETS Set-2 optic. EUV characterization of the Set-2 optic is performed using the EUV phase-shifting point diffraction interferometer (PS/PDI) installed on an undulator beamline at Lawrence Berkeley National Laboratory’s Advanced Light Source. This is the same interferometer previously used for the at-wavelength characterization and alignment of the ETS Set-1 optic. In addition to the PS/PDI-based full-field wave front characterization, we also present wave front measurements performed with lateral shearing interferometry, the chromatic dependence of the wave front error, and the system-level pupil-dependent spectral-bandpass characteristics of the optic; the latter two properties are only measurable using at-wavelength interferometry.
No takes yet. Share an insight, caveat, or question.
Naulleau et al. (2001) studied this question.
Synapse has enriched 4 closely related papers on similar clinical questions. Consider them for comparative context: