Answers are grounded in this paper and cite their sources.
This paper presents an optimized T-type resonant photoacoustic (PA) cell for methane (CH4) gas detection. The noise transmission coefficients and PA field distributions of the T-type resonant PA cell have been evaluated using the finite element method and thermoviscous acoustic theory. The optimized T-type resonant PA cell, together with a near-infrared (NIR) distributed feedback (DFB) laser source, a high-speed spectrometer and a fiber-optic acoustic sensor constitutes a PAS system for CH4 detection. The sensitivity is measured to be 1.8 pm/ppm and a minimum detectable limit (MDL) of 9 parts per billion (ppb) can be achieved with an averaging time of 500 s. The optimized T-type longitudinal resonant PA cell features of high PA cell constant, fast response time and simple manufacturing process.
No takes yet. Share an insight, caveat, or question.
Gong et al. (2020) studied this question.