Answers are grounded in this paper and cite their sources.
A silicon cantilever-enhanced photoacoustic spectroscopy (PAS)-based trace gas analyzer with low gas consumption is presented. A silicon cantilever-based fiber-optic Fabry–Perot (F–P) interferometric acoustic sensor with a compact structure and high sensitivity is designed for photoacoustic signal detection. The non-resonant photoacoustic cell (PAC) is a cylindrical copper tube with a volume of 0.56 mL. A near-infrared laser with a center wavelength of 1532.83 nm amplified using an erbium-doped fiber application amplifier is used as the excitation light. The wavelength modulation spectroscopy (WMS) technique is employed in the present work for second-harmonic photoacoustic signal detection. The experimental results show that the minimum detection limit of C 2 H 2 is 199.8 parts per trillion (ppt) with an average time of 60 s. The normalized noise equivalent absorption coefficient is calculated as 1.72 × 10 –9 cm –1 W/Hz 1/2 . Furthermore, the proposed silicon cantilever-enhanced non-resonant PAS-based gas analyzer can not only analyze the gas concentration in a closed small-capacity PAC with low gas consumption but also detect target gas leakage in real time at a long distance.
No takes yet. Share an insight, caveat, or question.
Guo et al. (2022) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: