This study describes the echocardiographic characteristics and measurement techniques for evaluating stent-mounted aortic bioprosthetic valves in the mitral position.
May guide echocardiographic assessment of mitral-position aortic bioprostheses; leaves open validation in larger cohorts.
Echocardiograms were performed on 20 clinically stable patients following mitral valve replacement with glutaraldehyde-preserved porcine aortic heterografts and three patients with antibiotic sterilized aortic homografts mounted in the mitral position. Such valves were evaluated in a test chamber at varied flow rates resulting in improved understanding of movements seen with the echocardiogram in vivo. The technique for recording the valvular stent and leaflets is described and a method for measuring several parameters is demonstrated. Initial diastolic slope averaged 2.4 +/- 0.5 cm/sec (range 1.9 to 3.3 cm/sec). Left ventricular outflow tract measured from the anterior portion of the stent to the interventricular septum averaged 1.5 +/- 0.5 cm at end-diastole and 1.3 +/- 0.6 cm at end-systole. Leaflet excursion averaged 1.5 +/- 0.3 cm (with a range from 1.0 to 2.1 cm). The ratio of internal to external stent diameters averaged 0.66 +/- 0.05 (with a range from 0.56 to 0.74).