Key result
Thoracic aortic stent-graft deployment using magnetic tracking and ultrasound guidance achieved a deployment error of 2.43 ± 0.69 mm in a porcine model.
Why the study?
Does a magnetic tracking system integrated with intraoperative ultrasound enable accurate thoracic aortic stent-graft deployment without ionizing radiation in preclinical models?
Does a magnetic tracking system integrated with intraoperative ultrasound enable accurate thoracic aortic stent-graft deployment without ionizing radiation in preclinical models?
A novel magnetic tracking system with intraoperative ultrasound enables accurate, radiation-free deployment of thoracic aortic stent-grafts in preclinical models.
We propose a system for thoracic aortic stent-graft deployment that employs a magnetic tracking system (MTS) and intraoperative ultrasound (US). A preoperative plan is first performed using a general public utilities-accelerated cardiac modeling method to determine the target position of the stent-graft. During the surgery, an MTS is employed to track sensors embedded in the catheter, cannula, and the US probe, while a fiducial landmark based registration is used to map the patient's coordinate to the image coordinate. The surgical target is tracked in real time via a calibrated intraoperative US image. Under the guidance of the MTS integrated with the real-time US images, the stent-graft can be deployed to the target position without the use of ionizing radiation. This navigation approach was validated using both phantom and animal studies. In the phantom study, we demonstrate a US calibration accuracy of 1.5 ± 0.47 mm, and a deployment error of 1.4 ± 0.16 mm. In the animal study, we performed experiments on five porcine subjects and recorded fiducial, target, and deployment errors of 2.5 ± 0.32, 4.2 ± 0.78, and 2.43 ± 0.69 mm, respectively. These results demonstrate that delivery and deployment of thoracic stent-graft under MTS-guided navigation using US imaging is feasible and appropriate for clinical application.
No takes yet. Share an insight, caveat, or question.
Luo et al. (2012) studied Thoracic aortic stent-graft deployment (n=5). Magnetic tracking system (MTS) and intraoperative ultrasound (US) guidance was evaluated on Deployment error in animal study (mm). Thoracic aortic stent-graft deployment using magnetic tracking and ultrasound guidance achieved a deployment error of 2.43 ± 0.69 mm in a porcine model.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: