Why the study?
To determine whether endovascular chemofiltration with an ionic device can reduce systemic exposure and off-target biodistribution of doxorubicin during hepatic intra-arterial chemotherapy.
Does endovascular chemofiltration reduce systemic exposure and off-target biodistribution of doxorubicin during hepatic intra-arterial chemotherapy in a preclinical pig model?
Does endovascular chemofiltration reduce systemic exposure and off-target biodistribution of doxorubicin during hepatic intra-arterial chemotherapy in a preclinical pig model?
Endovascular chemofiltration significantly reduces systemic exposure and off-target cardiac accumulation of doxorubicin during hepatic intra-arterial chemotherapy in a preclinical model.
Hypothesis-generating for chemofiltration to limit doxorubicin cardiotoxicity; human trials needed before clinical translation.
Purpose To determine if endovascular chemofiltration with an ionic device (ChemoFilter [CF]) can be used to reduce systemic exposure and off-target biodistribution of doxorubicin (DOX) during hepatic intra-arterial chemotherapy (IAC) in a preclinical model. Materials and Methods Hepatic IAC infusions were performed in six pigs with normal livers. Animals underwent two 10-minute intra-arterial infusions of DOX (200 mg) into the common hepatic artery. Both the treatment group and the control group received initial IAC at 0 minutes and a second dose at 200 minutes. Prior to the second dose, CF devices were deployed in and adjacent to the hepatic venous outflow tract of treatment animals. Systemic exposure to DOX was monitored via blood samples taken during IAC procedures. After euthanasia, organ tissue DOX concentrations were analyzed. Alterations in systemic DOX exposure and biodistribution were compared by using one-tailed t tests. Results CF devices were well tolerated, and no hemodynamic, thrombotic, or immunologic complications were observed. Animals treated with a CF device had a significant reduction in systemic exposure when compared with systemic exposure in the control group (P <.009). Treatment with a CF device caused a significant decrease in peak DOX concentration (31%, P <.01) and increased the time to maximum concentration (P <.03). Tissue analysis was used to confirm significant reduction in DOX accumulation in the heart and kidneys (P <.001 and P <.022, respectively). Mean tissue concentrations in the heart, kidneys, and liver of animals treated with CF compared with those in control animals were 14.2 μg/g ± 1.9 (standard deviation) versus 26.0 μg/g ± 1.8, 46.4 μg/g ± 4.6 versus 172.6 μg/g ± 40.2, and 217.0 μg/g ± 5.1 versus 236.8 μg/g ± 9.0, respectively. Fluorescence imaging was used to confirm in vivo DOX binding to CF devices. Conclusion Reduced systemic exposure and heart bioaccumulation of DOX during local-regional chemotherapy to the liver can be achieved through in situ adsorption by minimally invasive image-guided CF devices.
No takes yet. Share an insight, caveat, or question.
Yee et al. (2019) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: