Answers are grounded in this paper and cite their sources.
Poloxamers are polyethylene-polypropylene glycol linear co-polymers. A simple matrix-assisted laser desorption/ionization mass spectrometry (MALDI-MS) method has been developed for the determination of the average molecular weight of poloxamers. The molecular mass of five standard poloxamers determined by MALDI closely corresponds to that specified by the manufacturers, and no mass distribution effects were observed. Quantitation of distributions based on the molecular mass envelope using electrospray (ES) ionization was unsuccessful. To overcome this problem, quantitation was based on fragment ions (m/z 45 and 59) which gave reproducible signals using a very high orifice voltage ( approximately 200 eV). Poloxamer concentrations were determined accurately with a good linear response using the standard addition method. We believe that the use of very small fragment ions for quantitation of polymers may become a widely applicable general technique.
No takes yet. Share an insight, caveat, or question.
Takáts et al. (2001) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: