Answers are grounded in this paper and cite their sources.
An amperometric immunosensor for IgG was developed by covalently immobilizing anti‐IgG on multiwall carbon nanotube‐embedded conducting polymer, poly‐5,2′ : 5′′,2′′‐terthiophene‐3′‐carboxylic acid (MWCNT/pTTCA). The MWCNT/pTTCA modified electrode was characterized by SEM, EIS, and XPS. A hydrazine‐labeled secondary antibody‐MWCNT conjugate (Hyd‐MWCNT‐Ab 2 ) was applied for detection. Hydrazine was used as a catalyst for the reduction of hydrogen peroxide, which was monitored at −0.3 V vs. Ag/AgCl. The calibration plots showed a linear range of 0.1–10 ng/mL with a detection limit of 0.084±0.004 ng/mL. The proposed immunosensor was evaluated for clinical applications in a rabbit serum sample.
No takes yet. Share an insight, caveat, or question.
Zhu et al. (2010) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: