Conventional multichannel superconducting quantum interference device (SQUID) systems require a SQUID read-out circuit for each channel, as well as many wires connecting each individual SQUID and feedback coil to the room temperature electronics. We present a simple time domain multiplexed read-out scheme which requires only a single SQUID read-out circuit that is successively switched between all the SQUIDs. By connecting all the SQUIDs and all the feedback coils in series, this time domain multiplexed system requires many fewer wires between the SQUIDs and the room temperature read-out circuit than other multichannel systems.
No takes yet. Share an insight, caveat, or question.
Mück et al. (2004) studied this question.