We have investigated the microstructural and electron transport properties of 45° step-edge Josephson junctions grown on MgO substrates and used them for the preparation of superconducting quantum interference device (SQUID) magnetometers intended for magnetoencephalography (MEG) measurement systems. The high-TcSQUID magnetometers also incorporate 16 mm multilayer superconducting flux transformers on the MgO substrates and demonstrate a magnetic field resolution of ~ 4 fT/√Hz at 77 K. Results are illustrated for the detection of auditory evoked magnetic responses of the human cortex and compared between high-TcSQUIDs and a commercial low-TcMEG system. Our results demonstrate that MEG systems can be upgraded using high-TcSQUIDs to make them independent of helium and more user-friendly, saving operating costs and leading to the widespread utilization of MEG systems in clinical practice and at universities.
No takes yet. Share an insight, caveat, or question.
Faley et al. (2012) studied this question.