In this paper, we study a motion-compensated approach for simultaneous reconstruction of image frames in a time sequence. We treat the frames in a sequence collectively as a single function of both space and time, and define a temporal prior to account for the temporal correlations in a sequence. This temporal prior is defined in a form of motion-compensation, aimed to follow the curved trajectories of the object motion through space-time. The image frames are then obtained through estimation using the expectation-maximization (EM) algorithm. The proposed algorithm was evaluated using the four-dimensional (4-D) gated mathematical cardiac-torso (gMCAT) D1.01 phantom to simulate gated single photon emission computed tomography (SPECT) perfusion imaging with Tc-99m-sestamibi. Our experimental results demonstrate that the use of motion compensation for reconstruction can lead to significant improvement in image quality and reconstruction accuracy.
No takes yet. Share an insight, caveat, or question.
Gravier et al. (2005) studied this question.
Synapse has enriched 2 closely related papers on similar clinical questions. Consider them for comparative context: