A sensor for monitoring total ion activity is described, and its performance as a wearable device for monitoring the total ion levels of sweat is evaluated. The sensor works by tracking changes in the Donnan potential generated across a Nafion® membrane. This cation‐exchange polymer was cast on a paper coated with carbon‐ink, making the platform elegantly simple. Analytical parameters during calibration in aqueous solution include a sensitivity of 56.3±1.0 mV/dec.a (Na+) and a standard deviation between standard electrode potentials of 5.3 mV (N=5) for first time use. By integrating a paper‐based pseudo‐reference electrode, a miniature disposable electrochemical cell (the “IonSens” device) was created and demonstrated as a wearable sensor. Potentiometric measurements estimating the total ion activities were validated against conductivity measurements. Recoveries of eleven raw sweat samples were determined to be 95.2±6.6 % (n=3). The perspiration conductivity profile of an athlete during exercise was monitored in real‐time and visualized on a mobile phone application connected via Bluetooth®. The excellent reproducibility of the electrode without any conditioning is noteworthy and lends itself to applications including – but not limited to – the monitoring of total ion activity in sweat.
No takes yet. Share an insight, caveat, or question.
Hoekstra et al. (2018) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: