PURPOSE: To induce irreversible gelation of poloxamer, a thermosensitive polymer hydrogel, by using a photoinitiator and ultraviolet (UV) irradiation and to verify the biocompatibility and use of poloxamer as an injectable intraocular lens (IOL) material. SETTING: Department of Ophthalmology, Seoul National University College of Medicine, Seoul Artificial Eye Center, Clinical Research Institute, Seoul National University Hospital, Seoul, Korea. METHODS: In 10 rabbits, endocapsular phacoemulsification was performed and a poloxamer-photoinitiator mixture was injected into the capsular bag through a small capsulorhexis site. In 1 eye, the capsulorhexis site was closed with a small plug and the entire eye was irradiated with UV light for 5 minutes. Postoperatively, poloxamer transparency and effect on the conjunctiva, cornea, iris, vitreous, and retina were observed. RESULTS: A mixture comprising 25% poloxamer and 0.01% photoinitiator produced a poloxamer that remained transparent in the lens capsule for up to 6 months. No inflammatory response or toxicity was observed in the conjunctiva, cornea, iris, vitreous, or retina. CONCLUSION: Poloxamer is a potentially suitable material for an injectable IOL. Further study is needed.