Standard platinum resistance thermometers that are constructed with transparent sheaths of fused silica or pyrex glass are subject to thermal losses arising from losses of visible and near-infrared radiation up (piping) and through these sheaths. This heat loss may introduce substantial errors in temperature determinations; e.g. 0.084, 0.015, and 0.001 1 °C at the Sb, Zn, and Sn fixed-temperature points for a fused-silica sheath, or 0.000 2 °C at the Zn and Sn points for a pyrex sheath. This effect has been investigated in detail for thermometers of various types at the Sb, Zn, and Sn points, and it has been shown that simple stem radiation-traps, consisting of a blackening or roughening of the smooth outer surface of the thermometer sheaths, will eliminate these errors and greatly improve the immersion characteristics of the thermometers.Suitable radiation-trapping should be incorporated on the stems of every transparent-sheathed standard platinum resistance thermometer; without strong evidence to the contrary, radiation loss errors of the orders of magnitude cited above should be suspected in reported measurements involving resistance thermometers having unprotected transparent fused-silica sheaths.