High permeability magnetic flux concentrators are used to enhance the sensitivity of an atomic magnetometer operating in the spin-exchange relaxation-free regime. The magnetometer uses a millimeter scale Rb87 vapor cell and either mu-metal or Mn–Zn ferrite flux concentrators. The measured sensitivity gives excellent agreement with calculations of thermal noise from the concentrator material. The mu-metal concentrators allow a sensitivity of 50fTHz−1∕2, limited by thermal current magnetic noise. The ferrite concentrators are limited by thermal magnetization noise at low frequencies, and reach a sensitivity of 10fTHz−1∕2 for frequencies above 125Hz.
No takes yet. Share an insight, caveat, or question.
Griffith et al. (2009) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: