Recent technical improvements and cost reductions in electromagnetic motion tracking systems invite their application to motion axis determination in the surgical setting. After evaluation of the accuracy of a state-of-the-art D.C. electromagnetic tracking system, which generates complete three-dimensional kinematic outputs from just a single receiver, we calculated screw displacement axes (SDA's) from its source data. The accuracy of SDA determination from such source data was evaluated for various rotational increment sizes around a revolute joint. A novel smoothing procedure, customized for this type of source data, was developed, enabling SDA detection from incremental rotations of less than 1 deg, at an accuracy appropriate for intra-operative measurement of human joint motion. Examples of SDA determination are given for motion tracking of a ball joint and of the elbow articulation.
No takes yet. Share an insight, caveat, or question.
Bottlang et al. (1998) studied this question.
Synapse has enriched 4 closely related papers on similar clinical questions. Consider them for comparative context: