We present an efficient, low-cost modulation transfer function (MTF) measurement approach, optimized for characterization of tunable micro-lenses; the MTF may easily be measured at a variety of different focal lengths. The approach uses a conventional optical microscope with an optimized approach for lens illumination and the measurement results have been correlated with a commercial MTF measurement system. Measurements on fixed-focus and tunable micro-lenses were performed; for the latter, resolution for lenses with back focal length of 11 mm was 55 lines/mm, decreasing to 40 lines/mm for a back focal length of 4 mm. In general, it was seen that performance was better for lenses with longer focal lengths.
No takes yet. Share an insight, caveat, or question.
Zhang et al. (2010) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: