Key result
PISA strongly correlates with reference methods for calculating EROA in aortic regurgitation.
Why the study?
Effective regurgitant orifice area can be calculated using the proximal isovelocity surface area method, but this approach had not been validated in aortic regurgitation.
Does the PISA method accurately calculate the effective regurgitant orifice area compared to reference methods in patients with isolated aortic regurgitation?
Observational (n=71)
Does the PISA method accurately calculate the effective regurgitant orifice area compared to reference methods in patients with isolated aortic regurgitation?
Effect estimate: r=0.90
p-value: p=< 0.0001
The PISA method provides reliable measurement of the effective regurgitant orifice in aortic regurgitation, except in patients with an obtuse flow convergence angle where it may underestimate severity.
OBJECTIVES: We sought to determine the reliability of the proximal isovelocity surface area (PISA) method for calculation of effective regurgitant orifice (ERO) of aortic regurgitation (AR). BACKGROUND: The ERO area can be calculated by the PISA method, but this method has not been validated in AR. METHODS: ERO calculation by the PISA method was undertaken prospectively in 71 consecutive patients with isolated AR and achieved in 64 and compared with two simultaneous reference methods (quantitative Doppler and quantitative two-dimensional echocardiography). In addition, this method was compared with angiography in 12 patients, with surgical assessment in 18 patients and with ventricular volumes in all patients. RESULTS: Good correlations between PISA and reference methods were obtained (both r=0.90, both p < 0.0001), but a trend toward underestimation of the ERO by the PISA method was noted (24+/-19 vs. 26+/-22 mm2 and 27+/-23 mm2, respectively, both p=0.04). However, this trend was confined to five patients with an obtuse flow convergence angle (>220 degrees), and on multivariate analysis this variable was the only independent determinant of underestimation of the ERO. In contrast, in 59 patients with a flat flow convergence (< or =220 degrees ), the PISA method, in comparison with reference methods, showed excellent correlations, with a narrow standard error of the estimate (r=0.95, SEE 5.4 mm2, and r=0.95, SEE 5.8 mm2; all p < 0.0001) and no trend toward underestimation (22+/-18 vs. 23+/-16 mm2, p=0.44, and vs. 23+/-18 mm2, p=0.34). CONCLUSIONS: In patients with AR, the PISA method can be used to measure the ERO with reasonable feasibility. Underestimation of the ERO by PISA may occur in patients with an obtuse flow convergence angle. However, in most patients with appropriate flow convergence, PISA provides reliable measurement of the ERO of AR.
No takes yet. Share an insight, caveat, or question.
Tribouilloy et al. (1998) conducted an observational in Isolated aortic regurgitation (n=71). Proximal isovelocity surface area (PISA) method vs. Quantitative Doppler and quantitative two-dimensional echocardiography was evaluated on Effective regurgitant orifice (ERO) area correlation with reference methods (r=0.90, p=< 0.0001). The proximal isovelocity surface area (PISA) method showed good correlation with reference methods for calculating the effective regurgitant orifice area in aortic regurgitation (r=0.90, p<0.0001).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: