BACKGROUND: An evaluation of SENTiFIT® 270 (Sentinel Diagnostics, Italy; Sysmex, Spain) analyser for the quantitation of faecal haemoglobin (f-Hb) was performed. METHODS: The analytical imprecision, linearity, carry over and f-Hb stability were determined. Evaluation of the diagnostic accuracy was performed on 487 patients. RESULTS: Within-run and between-run imprecision ranged 1.7%-5.1% and 3.8%-6.2%, respectively. Linearity studies revealed a mean recovery of 101.1% (standard deviation, 6.7%) for all dilutions. No carry over was detected below 7650 μg Hb/g faeces. Decay of f-Hb in refrigerated samples ranged 0.2%-0.5% per day. f-Hb in patients with advanced colorectal neoplasia (ACRN) (colorectal cancer [CRC] plus advanced adenoma [AA]) were significantly higher than from those with a normal colonoscopy. Sensitivity for ACRN at f-Hb cutoffs from 10 to 60 μg Hb/g faeces ranged from 28.9% (95% confidence interval [CI], 21.7%-37.2%) to 46.5% (95% CI, 38.1%-55%), the specificity ranged from 85% (95% CI, 82.3%-87.3%) to 93.2% (95% CI, 91.2%-94.8%), positive predictive values for detecting CRC and AA ranged from 11.6% (95% CI, 7.6%-17.2%) to 20.6% (95% CI, 13.3%-30.3%) and from 34.7% (95% CI, 28.1%-42%) to 42.3% (95% CI, 32.4%-52.7%), respectively, and the negative predictive value for ACRN ranged from 90.2% (95% CI, 87.9%-92.2%) to 88.4% (95% CI, 86%-90.4%). Using two samples per patient sensitivity increased with a slight decrease in specificity. CONCLUSIONS: The analytical and clinical performances of SENTiFIT assay demonstrate a specific and accurate test for detecting ACRN in symptomatic patients and those undergoing surveillance.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Augé et al. (2017) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: