Answers are grounded in this paper and cite their sources.
BACKGROUND: The sensitivity of cobalt (Co) and chromium (Cr) ion-levels in detecting poorly performing metal-on-metal hip implants is low. This study proposes that serial changes in ion-levels are a more accurate marker of arthroplasties at risk. METHODS: Serial metal ion-levels and implant data of 285 patients with ASR resurfacing or replacement were studied. Patient and implant characteristics were analysed using univariate and multivariate analyses. RESULTS: 111 (39%) had revision surgery. Time since index surgery (p<0.001), acetabular inclination (p<0.001), their interaction (p<0.001) and femoral head size (p = 0.01) were significant variables. Head size (≤51 mm) had lower Co and Cr levels (p = 0.01). Cr/Co showed marginal decrease over time in the unrevised group and no decrease prerevision. CONCLUSIONS: Repeated measurement of ion-levels were higher in the revision group suggesting that serial measurements rather than absolute values may have a role to play in predicting implant failure.
No takes yet. Share an insight, caveat, or question.
McHugh et al. (2017) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: