OBJECTIVE: The purpose of this paper is to demonstrate that a new algorithm for estimating arterial oxygen saturation can be effective even with data corrupted by motion artifacts (MAs). METHODS: OxiMA, an algorithm based on the time-frequency components of a photoplethysmogram (PPG), was evaluated using 22-min datasets recorded from 10 subjects during voluntarily-induced hypoxia, with and without subject-induced MAs. A Nellcor OxiMax transmission sensor was used to collect an analog PPG while reference oxygen saturation and pulse rate (PR) were collected simultaneously from an FDA-approved Masimo SET Radical RDS-1 pulse oximeter. RESULTS: The performance of our approach was determined by computing the mean relative error between the PR/oxygen saturation estimated by OxiMA and the reference Masimo oximeter. The average estimation error using OxiMA was 3 beats/min for PR and 3.24% for oxygen saturation, respectively. CONCLUSION: The results show that OxiMA has great potential for improving the accuracy of PR and oxygen saturation estimation during MAs. SIGNIFICANCE: This is the first study to demonstrate the feasibility of a reconstruction algorithm to improve oxygen saturation estimates on a dataset with MAs and concomitant hypoxia.