Cyclic voltammetric measurements were made using well‐characterized microcrystalline boron‐doped diamond thin‐film electrodes to test the material's responsiveness for ferrocene as a function of scan rate, solvent, and electrolyte composition. Apparent heterogeneous electron transfer rate constants, k ° app , of 0.042±0.015, 0.048±0.015, and 0.008±0.002 cm/s were observed in 0.1 M NaClO 4 /CH 3 CN, 0.1 M TBAClO 4 /CH 3 CN, and 0.1M TBAClO 4 /CH 2 Cl 2 , respectively. These rate constants, obtained using electrodes without any type of pretreatment, are similar to those observed for freshly polished glassy carbon. The results demonstrate that boron‐doped diamond is a viable material for the electrochemical analysis of nonaqueous analytes.