Answers are grounded in this paper and cite their sources.
We have developed a multi-region spherical Monte Carlo (MC) model to simulate the dynamic changes in light intensity measured during laser interstitial thermal therapy (LITT). Model predictions were validated experimentally in tissue-simulating albumen phantoms with well-characterized optical properties that vary dynamically with LITT in a way similar to tissue. For long treatments (2.5 W, approximately 1800 s), the transient light intensity changes demonstrated better qualitative agreement with a three-region MC model (with an inner layer of fully coagulated optical properties, a middle layer of partially coagulated properties and an outer region of native properties); for short treatments (4 W, approximately 240 s), better qualitative agreement was seen with a two-region MC model (with an inner layer of fully coagulated properties and outer region of native properties). These differences were attributed to differences in coagulation formation during low- and high-powered heating regimes, respectively. At the end of heating, a three-region coagulation zone was observed for both heating schemes. Quantitatively, final light intensity changes at the end of heating were compared with changes predicted by both two- and three-region MC for the same experimentally measured coagulation size and found to agree within approximately 30% for both models. The developed MC model helps lend insight into the nature of thermal coagulation events occurring for low and high power LITT irradiation schemes.
No takes yet. Share an insight, caveat, or question.
Chin et al. (2003) studied this question.