This paper reports results, experimental and analytical, of the nonlinear behavior of a shim coil in the presence of an HTS coil assembled with double-pancake (DP) HTS-tape coils. The experimental results are from: 1) operation of a 700 MHz LTS/HTS NMR magnet (LH700) consisting of a 600 MHz LTS NMR magnet (L600) equipped with superconducting shim coils and a 100 MHz DP-assembled HTS insert (H100) and; 2) an experiment with a room-temperature (RT) Z1 shim coil coupled to a 50 MHz DP-as-sembled HTS insert (H50). A field mapping theory with a harmonic analysis is applied to interpret both results. Based on experimental results and analyses, we conclude that the screening-current-induced field (SCF) generated by a DP-assembled HTS insert is primarily responsible for the nonlinear behavior, including strength degradation, of a shim coil.