An improved version of our magnetic suspension densimeter has been developed for use with samples of aqueous systems as small as 100 μl. The instrument’s sensitivity to a change in density of the solution under test (measured as a voltage shift) is 0.1 mV≊4.5×10−6 g ml−1, with a base-line drift less than 9 μV/h. The nonlinearity of the densimeter’s response is less than 5 ppm over its calibration range. In this paper, we describe the new magnetic suspension circuit, discuss the precision limits of the instrument, and present the results of measurements made on a tetracarboxylate sequestering agent known as BAPTA which we use as a calcium-ion indicator.