A new approach to detecting capillary electrophoresis (CE) eluent components by interfacing CE with a surface-enhanced Raman scattering (SERS) system is described. In this approach, CE-based separation of a mixture of trans-1,2-bis(4-pyridyl)ethylene and N,N-dimethyl-4-nitrosoaniline has been detected by SERS in a postcolumn geometry. The retention time obtained from SERS corresponds well with that from conventional UV-visible detection. Meanwhile, CE eluants are identified by their characteristic vibrational spectra, demonstrating the validity of SERS as a structure-specific detection method for CE. In addition, the ability to monitor SERS intensity changes at molecule-specific frequencies makes selective detection of individual analytes possible, even when separation is incomplete. Finally, CE-SERS is evaluated for separation of amino acids (tyrosine and tryptophan) and environmental pollutants (chlorophenol mixtures).
No takes yet. Share an insight, caveat, or question.
He et al. (2000) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: