A transportable GC x GC instrument is under development for on-site applications that would benefit from the enhanced resolution and powers of detection, which can be achieved by this method. In the present study, a low-resource GC x GC instrument using an electrically heated and liquid-cooled single-stage thermal modulator that requires no cryogenic materials is evaluated. The instrument also uses at-column heating, thus eliminating the need for a convection oven to house the two columns. The stainless-steel modulator tube is coated with PDMS, which can be heated to 350 degrees C for sample injection into the second-dimension column. The modulator is cooled to -30 degrees C by a 100 mL/min flow of PEG by means of a commercial liquid chiller and a small recirculating pump. Resistive heating of the modulator tube is provided by a programmable power supply, which uses a voltage program that results in increasing modulator temperature during an analysis. This, together with more rapid cooling by the use of a liquid cooling medium, results in reduced solute breakthrough following each heating cycle as the modulator cools to a temperature where quantitative trapping resumes. As a result, modulated peak widths at half-height of less than 40 ms are observed. Design and performance details are presented along with chromatograms of gasoline and an essential oil sample.
No takes yet. Share an insight, caveat, or question.
Libardoni et al. (2006) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: