Answers are grounded in this paper and cite their sources.
The coupling of glass capillary supercritical fluid chromatography to a high temperature GC/MS system via a micrometeradjustable glass capillary interface including an integrated pressure restrictor is described. With this coupling device, both complementary capillary chromatographic methods retain their full independence and flexibility. It is shown that in supercritical fluid chromatography glass fulfils all the requirements of a suitable support material. The preparation of narrow bore glass capillary columns (0.06 mm i.d.), coated with chemically bonded and crosslinked fluids with a film thickness of about 0.6 μm, which exploit the merits of OHterminated polysiloxanes as stationary phases is discussed. The application of glass capillary SFC is demonstrated and compared with examples given in the literature.
No takes yet. Share an insight, caveat, or question.
Blum et al. (1988) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: