We modified the Incstar Cyclo Trac SP kit to enable its use with dry blood-spots on filter paper. The recovery ranged from 92 to 106%. Dilution studies have shown excellent linearity and parallelism throughout the range of the assay. Precision is demonstrated by within-assay CV's of 6.6 and 4.3% at 96 and 342 micrograms/L respectively and between-assay CV's of 9.1 and 7.0% at 138 and 506 micrograms/L respectively. A comparison study (n = 209) with whole blood assay gave a correlation coefficient of 0.97, a slope of 1.04, and an intercept of 13.2. Whole blood and dry blood-spot cyclosporine assays on heart, kidney, liver, and lung transplants were also compared.