A detector for determining the intensity and duration of kiloampere pulsed proton beams is described. Proton intensities are inferred from measurements of prompt gamma rays of 6 to 7 MeV from the 19F(p, αγ)16O reaction. The thick-target yield for this reaction on a Teflon target was measured for 0.5-to 2.8-MeV protons. Two different size detectors were calibrated absolutely using the continuous proton beam from a Van de Graaff accelerator. An absolute measurement of a kiloampere pulsed proton beam is described, and the measured current is in agreement with an independent determination of the total proton intensity by nuclear activation.
No takes yet. Share an insight, caveat, or question.
Young et al. (1981) studied this question.