We consider microwave tomography (MWT) where the imaging region is surrounded by an electrically conducting surface. This surface acts as both a shield from outside interference, holding tank for any possible matching media, and, in certain cases, serves to enhance the performance of electromagnetic (EM) inversion algorithms. For the 2D transverse magnetic (TM) case and where the surface consists of a perfect electrical conductor (PEC) in the shape of a circular cylinder, we formulate an appropriate Greens function which is amenable to implementation in the existing EM inversion codes. We utilize this Greens function in the multiplicative-regularized contrast source inversion (MR-CSI) method. Several different synthetic examples are used to test the performance of the inversion when the PEC surface is present and the results show that in many cases, the tomographic image is significantly improved. The reasons for the improved inversion results are an area of active research, but are likely to be due to the increased interrogation energy deposited into the imaging region. Results are also shown which demonstrate the problems which may arise if the unbounded domain Greens function is used in an MWT system that utilizes a matching medium of finite extent—a problem which is overcome by the inclusion of a PEC surface on the exterior of the MWT system.