PURPOSE: To determine in vitro, using a translational research approach before realizing the procedure in a patient with iatrogenic aphakia and partial aniridia, whether suturing a trifocal intraocular lens (IOL) to an artificial iris degrades the IOL's optical quality. METHODS: Optical quality was analyzed by measuring the modulation transfer function (MTF) at a 3-mm aperture and at 50 and 100 lp/mm spatial frequencies. The FineVision Pod F GF IOL (PhysIOL) was assessed in two powers: two +20.00 diopters (D) (20A and 20B IOLs) and two +30.00 D (30A and 30B IOLs). The IOLs' decentration in relation to the artificial iris's center was evaluated. The laboratory results provided empirical evidence in the informed consent for surgical intervention in a patient with iatrogenic aphakia and iris defect in one eye. Clinical results were measured using the parameter of corrected distance visual acuity plus a patient self-assessment of the cosmetic appearance of the operated eye. RESULTS: The 20A and 20B IOLs demonstrated a mean MTF reduction of up to 1.1%, whereas the 30A and 30B IOLs showed a decrease of up to 5.2% for both spatial frequencies. All lenses showed good centration levels. In the clinical case, the patient showed corrected distance visual acuity, distance-corrected near visual acuity, and distance-corrected intermediate visual acuity of 0.20, 0.20, and 0.22 logMAR, respectively. The patient was satisfied with the cosmetic outcome. CONCLUSIONS: .
No takes yet. Share an insight, caveat, or question.
A 2022 study studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: